PerkinElmer Opera Phenix - Specifications

General information

Microscope type: Spinning disk confocal and widefield
Purpose: High-throughput and high-content imaging of multi-well plates will full automation
FPbase link: https://www.fpbase.org/microscope/rBKJsuQSCGgSnZuxcGRYSh/

Equipment information

Location: JSCBB C350C

Serial number: 1400L18161
CU Tag: UCB 214288 RC16

Sample types

Fixed Cells Yes
Live Cells Yes

Software

  • Harmony v5.6

Environmental control

  • CO2, Humidity, Temperature

Laser lines

  • 375 nm or 425 nm
  • 488 nm
  • 561 nm
  • 640 nm

Filters

Emission
Example fluorophores
Center (nm) Width (nm) Min (nm) Max (nm)
457.5 45 435 480 DAPI, BFP, CellMask Blue
472.5 15 465 480 CFP
475 80 435 515 CFP/YFP FRET: Acceptor Only
492.5 115 435 550 DAPI, BFP, CFP, CellMask Blue
497.5 65 465 530 DAPI, BFP, CFP, CellMask Blue
515 30 530 500 Alexa 488, GFP
525 50 500 550 GFP, Alexa 488, FITC, CellTracker Green, YFP
532.5 35 515 550 YFP
583.5 25 571 596 Alexa 568, Cy3, TRITC, dsRed
600 60 570 630 TRITC, Texas Red, dsRed, Alexa 568, Alexa 594, mCherry
617.5 25 605 630 Alexa 594, mCherry
665 30 650 680 Alexa 647, Cy5, CellMask Deep Red
705 30 690 720 Cy5.5, Draq5
705 110 650 760 Cy5, Alexa 647, Draq5

Please note that these emission filters are tied to specific laser lines/cameras, indicated by color above.

  • 375 nm or 425 nm laser → 457.5 nm, 472.5 nm, or 475 nm filter → Camera 1
  • 488 nm laser → 492.5 nm, 497.5 nm, 515 nm, 525 nm, 532.5 nm filter → Camera 2
  • 561 nm laser → 583.5 nm, 600 nm, or 617.5 nm filter → Camera 3
  • 640 nm laser → 665 nm or either 705 nm filter → Camera 4

Objectives

Magnification NA Immersion medium Field of view (mm) Effective xy resolution (µm),
binning = 1*
Effective xy resolution (µm),
binning = 2*
Recommended minimum distance between planes (µm)**
5X 0.16 Air 2.58 x 2.58 2.6 5.3 26.5
10X 0.3 Air 1.29 x 1.29 1.3 2.6 7.4
20X 0.8 Air 0.65 x 0.65 0.66 1.32 1.2
20X 1.0 Water 0.65 x 0.65 0.66 1.32 0.8
40X 1.1 Water 0.32 x 0.32 0.33 0.66 0.5

*Calculated at 520 nm. Optical resolution = 0.61 ʎemission / NA.

**Calculated as "Depth of focus / 2.2" (according to Nyquist's theorem). Please note that using smaller distances will not improve the results.

Detectors

  • Andor Zyla x4 sCMOS camera
  • 16-bit camera: pixel intensity values from 0 - 65,535
  • Camera ROI: 2160 x 2160 pixels

Instrument diagram

Opera_Layout_New.png