Acquiring a single image
Ifprotocolacquiringexplainsanhow to take a simple image with just a single frame and a singlechannel,channel.follow Acquiring a single image.- If acquiring multiple channels, timelapse, or z-stack images, follow Using the ND Acquisition
tool.
This
panel.
Acquiring a single image
- Select the desired optical configuration by clicking on the button. You will hear the microscope change filter cubes. Wait until the sound stops before proceeding.
- Check that the optical configuration settings are correct.
- Click on the icon. The microscope will take an image.
- Save the image by selecting File > Save As.
Using the ND Acquisition tool
Basic instructions:
Check that the channels you want to image has been set up correctly by following the instructionsChanging optical configurationsandRefining the perfect focus. Test that these settings give you the correct image byusing live viewand cycling through each channel.For each acquisition type (i.e. timelapse, z-stack, channels) you wish to acquire, make sure that the corresponding tab has been checked. Every other tab should be unchecked.Check that the settings in each acquisition type has been set correctly.To save the image to file, make sure the box next to “Save to File” is checked. Change the output path and filename as appropriate.We suggest:Output directory: Save your images under the folder that corresponds to your lab PI. If the folder doesn’t already exist, you can create it. Create a new subfolder with your name.Filename: Label images with the acquisition date and a short description, e.g. “20210816 HeLa Treated”.Write down all settings with the same heading in your lab notebook.
Acquire the image(s) by clicking on “Run now”.
Acquiring multiple fluorescence channels
First, make sure that the optical configuration has been set correctly for each channel that you intend to acquire by following the instructions underChanging optical configurations.Follow the protocol for using live view or acquiring a single image to change the optical settings.Alternatively, if you already know the correct settings (e.g. from a previous experiment), click on each optical configuration and adjust the setting. Make sure to click on thebutton to save the settings on each channel.Check that the optical configurations are as you expect by cycling through the channels WITHOUT turning on the acquisition. Making sure you have the correct settings now will save you time later.
Set up the ND Acquisition:Click on the checkbox next to the λ tab.If there are optical configurations in the list below, click on thebutton to clear themAdd each channel to the acquisition by:Clicking on thebuttonClick on the … button next to the channel name under the “Opt. Conf.” column. Select the optical configuration that you want to capture, then click OK.Check that the channel has a check mark next to it (you can skip a channel by clearing the check mark).Channels will be acquired from top to bottom. To change the order, select the channel and use the arrows to move them up or down.
If you are using the perfect focus system, make sure that the Use PFS button is checked
Setting up a Time-Lapse movie
In the ND Acquisition panel, check the box next to the Time tab.If there are time-lapse settings, click on thebutton to clear themAdd a new Phase to the acquisition by clicking on thebuttonEnter your values for the Interval, Duration and Loops:Interval = the amount of time between each acquisitionDuration = the total length of time for the imagingLoops = the number of images that will be acquired during the time-lapseNote:Only 2 out of 3 settings need to be entered, and the remaining setting will be calculated. For example, if you enter the Interval and the Duration, the number of loops will be automatically determined.
If you are using the perfect focus system, make sure that the Use PFS button is checkedCheck the box next to “Close Active Shutter when idle”. This turns off the excitation light between acquisitions so that you do not damage your sample.
Setting up a Z-stack
In the ND Acquisition panel, check the box next to the Z tab.Select one of the three buttons to set up a Z-stack.
Top and Bottom
Note: This option works best when not using PFS
Raise the objective to the top of your sample and select Top to set the maximum z valueLower the objective to the bottom of your sample and select Bottom to set the minimum z valueSet the z Step size (in µm) or the number of Steps. The optimal z Step size is indicated to the right of the Step value (e.g., 0.6 µm in the image above). You can either manually enter this value or click on the < button to set this height.
Define the Middle and move equal distance above and below
Note: This option works best with Perfect Focus
Raise the objective to the middle focal plane of your sampleSelect “Home” then “Relative”Define two out of the following three parameters: Step size, number of Steps or RangeNote:The optimal z Step size is indicated to the right of the of the Step value (0.6 µm in the image above). You can either manually enter this value or click on the button to set this height.
Define a 0 position and move unequal distances above and below
This option works best with Perfect Focus
Raise the objective a focal plane within your sampleSelect “Home” then “Relative”Define the distance Below and Above the defined plane to be imagedDefine either the Step size or the number of StepsNote: The optimal z Step size is indicated to the right of the of the Step value (0.6 µm in the image above). You can either manually enter this value or click on the button to set this height.
Setting up a Multipoint Acquisition
In the ND Acquisition panel, check the box for Multipoint Acquisition (XY).If there are positions listed, click on thebutton to clear themYou can choose to include the Z position by selecting the “Include Z” boxIf you arenotusing the PFS, it is recommended that you include zIf you are using the PFS, you typically do not include z
Move the stage to a desired XYZ positionAdd the position to the acquisition by clicking on thebuttonIf you are using the perfect focus system, make sure that the Use PFS button is checkedBe sure to select “Close Active Shutter during Stage Movement”. This turns off the excitation light between positions so that you do not damage your sample.