Nikon N-STORM - Specifications
General information
Microscope type: TIRF and super resolution
Purpose: STORM- or PALM-based super resolution, TIRF, HiLo, or FRAP
FPbase link: https://www.fpbase.org/microscope/p8tJqsbQSCuVeuQvhbrAM3/
Equipment information
Location: JSCBB C350B
Sample types
| Fixed Cells | Yes |
| Live Cells | Yes |
Software
- NIS Elements 6.2
Base
- Nikon Ti2
Environmental control
- CO2, Humidity, Temperature
Laser lines
TIRF/STORM laser box
- 405 nm
- 488 nm
- 561 nm
- 647 nm
FRAP/LUN4 laser box
- 405 nm
- 488 nm
- 561 nm
- 640 nm
Filters
For light coming from FRAP/LUN4 laser box:
| Filter (Turret-Hi) | Excitation (nm) | Emission (nm) | Dichroic mirror (nm) | Purpose |
| LF405-C (405 nm TIRF) | 370 – 410 | 429 – 474 | 417 – 1200 | Calibrating the galvo for FRAP |
| Continuous STORM |
387 – 417 483 – 494 557 – 570 636 – 661 |
422 – 478 502 – 549 581 – 625 674 – 786 |
420 – 481 497 – 553 575 – 628 667 – 792 |
Reflecting any of the lasers to the sample for FRAP |
| Filter (Turret-Lo) | Excitation (nm) | Emission (nm) | Dichroic mirror (nm) | Purpose |
| Empty |
— |
— |
— |
Allow light from sample to reach the camera |
For light coming from the TIRF/STORM laser box:
| Filter (Turret-Lo) | Dichroic mirror (nm) | Purpose |
| 86000v2 - Sedat Quad |
444 – 479 499 – 541 567 – 625 649 – ∞ |
Reflecting any of the lasers to the sample for TIRF/STORM/laser excitation |
| Filter (Turret-Hi) | Dichroic mirror (nm) | Purpose |
| Empty |
— |
Allow light to reach the sample or the camera |
For light coming from the Sola II LED box:
| Filter (Turret-Lo) | Excitation (nm) | Emission (nm) | Dichroic mirror (nm) | Purpose |
| DAPI | 340 – 380 | 435 – 485 | 400 – ∞ | Image DAPI, BFP |
| FITC |
465 – 495 |
515 – 555 |
505 – ∞ |
Image EGFP, ATTO 488, FITC |
| TRITC |
528 – 553 |
577 – 630 |
565 – ∞ |
Image RFP, ATTO 565, TRITC |
| Cy5 |
590 – 650 |
663 – 738 |
660 – ∞ |
Image iRFP, Alexa 647, Cy5 |
| Filter (Turret-Hi) | Excitation (nm) | Emission (nm) | Dichroic mirror (nm) | Purpose |
| Empty |
— |
— |
— |
Allow light to reach the sample or the camera |
Objectives
| Magnification | NA | Immersion medium |
| 10X | 0.45 | Air |
| 20X | 0.75 | Air |
| 60X TIRF | 1.49 | Oil |
| 100X TIRF | 1.49 | Oil |
Detectors
- Two Andor iXon 897 (EMCCD) cameras equipped with TwinCam for simultaneous imaging of two colors
- 16-bit detector: pixel intensities range from 0- 65,535
- Full sensor: 512 x 512 pixels
- Field size: (number of pixels in x) x (pixel size in x)
- 10x objective = 819.2 μm x 819.2 μm
- Pixel size in x,y: (16.0 μm / objective magnification) x binning2
- 10x objective at no binning = 1.60 μm x 1.60 μm in x,y
- 10x objective at 2x2 binning = 6.4 μm x 6.4 μm in x,y
- One Hamamatsu Orca Quest 2 (qCMOS) camera equipped with a DoubleHelix add-on for the addition of phase masks
- 16-bit detector: pixel intensities range from 0 - 65,535
- Full sensor: 4096 x 2304 pixels
- Field size: (number of pixels in x) x (pixel size in x)
- 10x objective = 1884.16 μm x 1059.84 μm
- Pixel size in x,y: (4.6 μm / objective magnification) x binning2
- 10x objective at no binning = 0.46 μm x 0.46 μm in x,y
- 10x objective at 2x2 binning = 1.84 μm x 1.84 μm in x,y
History
| Dates | Instrument change |
| 02-12-2025 – 02-11-2026 | Ran on software NIS Elements 6.10.01 |
| Before 02-12-2025 | Ran on software NIS Elements 5 |
No comments to display