For questions, comments or concerns please contact Imaging
Skip to main content

Nikon N-STORM - Specifications

General information

Microscope type: TIRF and super resolution
Purpose: STORM- or PALM-based super resolution, TIRF, HiLo, or FRAP
FPbase link: https://www.fpbase.org/microscope/p8tJqsbQSCuVeuQvhbrAM3/

Equipment information

Location: JSCBB C350B

Sample types

Fixed Cells Yes
Live Cells Yes

Software

  • NIS Elements 6.2

Base

  • Nikon Ti2

Environmental control

  • CO2, Humidity, Temperature

Laser lines

TIRF/STORM laser box

  • 405 nm
  • 488 nm
  • 561 nm
  • 647 nm

FRAP/LUN4 laser box

  • 405 nm
  • 488 nm
  • 561 nm
  • 640 nm

Filters

For light coming from FRAP/LUN4 laser box:

Filter (Turret-Hi) Excitation (nm) Emission (nm) Dichroic mirror (nm) Purpose
LF405-C (405 nm TIRF) 370 – 410 429 – 474 417 – 1200 Calibrating the galvo for FRAP
Continuous STORM

387 – 417

483 – 494

557 – 570

636 – 661

422 – 478

502 – 549

581 – 625

674 – 786

420 – 481

497 – 553

575 – 628

667 – 792

Reflecting any of the lasers to the sample for FRAP
Filter (Turret-Lo) Excitation (nm) Emission (nm) Dichroic mirror (nm) Purpose
Empty

Allow light from sample to reach the camera

 

For light coming from the TIRF/STORM laser box:

Filter (Turret-Lo) Dichroic mirror (nm) Purpose
86000v2 - Sedat Quad

444 – 479

499 – 541

567 – 625

649 – ∞

Reflecting any of the lasers to the sample for TIRF/STORM/laser excitation
Filter (Turret-Hi) Dichroic mirror (nm) Purpose
Empty

Allow light to reach the sample or the camera

 

For light coming from the Sola II LED box:

Filter (Turret-Lo) Excitation (nm) Emission (nm) Dichroic mirror (nm) Purpose
DAPI 340 – 380 435 – 485 400 – ∞ Image DAPI, BFP
FITC

465 – 495

515 – 555

505 – ∞

Image EGFP, ATTO 488, FITC
TRITC

528 – 553

577 – 630

565 – ∞

Image RFP, ATTO 565, TRITC
Cy5

590 – 650

663 – 738

660 – ∞

Image iRFP, Alexa 647, Cy5
Filter (Turret-Hi) Excitation (nm) Emission (nm) Dichroic mirror (nm) Purpose
Empty

Allow light to reach the sample or the camera

Objectives

Magnification NA Immersion medium
10X 0.45 Air
20X 0.75 Air
60X TIRF 1.49 Oil
100X TIRF 1.49 Oil

Detectors

  • Two Andor iXon 897 (EMCCD) cameras equipped with TwinCam for simultaneous imaging of two colors
    • 16-bit detector: pixel intensities range from 0- 65,535
    • Full sensor: 512 x 512 pixels
    • Field size: (number of pixels in x) x (pixel size in x)
      • 10x objective = 819.2 μm x 819.2 μm
    • Pixel size in x,y: (16.0 μm / objective magnification) x binning2
      • 10x objective at no binning = 1.60 μm x 1.60 μm in x,y
      • 10x objective at 2x2 binning = 6.4 μm x 6.4 μm in x,y
  • One Hamamatsu Orca Quest 2 (qCMOS) camera equipped with a DoubleHelix add-on for the addition of phase masks
    • 16-bit detector: pixel intensities range from 0 - 65,535
    • Full sensor: 4096 x 2304 pixels
    • Field size: (number of pixels in x) x (pixel size in x)
      • 10x objective = 1884.16 μm x 1059.84 μm
    • Pixel size in x,y: (4.6 μm / objective magnification) x binning2
      • 10x objective at no binning = 0.46 μm x 0.46 μm in x,y
      • 10x objective at 2x2 binning = 1.84 μm x 1.84 μm in x,y

History

Dates Instrument change
02-12-2025 – 02-11-2026 Ran on software NIS Elements 6.10.01
Before 02-12-2025 Ran on software NIS Elements 5