PerkinElmer Opera Phenix

PerkinElmer Opera Phenix Specifications

General information

Microscope type: Spinning disk confocal and widefield
Purpose: High-throughput and high-content imaging of multi-well plates will full automation
FPbase link: https://www.fpbase.org/microscope/rBKJsuQSCGgSnZuxcGRYSh/

Equipment information

Location: JSCBB C350C

Serial number: 1400L18161
CU Tag: UCB 214288 RC16

Sample types

Fixed Cells Yes
Live Cells Yes

Software

Environmental control

Laser lines

Filters

Emission

Example fluorophores

Vendor

Model number
Center (nm) Width (nm) Min (nm) Max (nm)
457.5 45 435 480 DAPI, BFP, CFP Semrock FF01-457/50
492.5 115 435 550 DAPI, BFP, CFP    
475 80 435 515 DAPI, BFP, CFP    
525 50 500 550 GFP, Alexa Fluor 488, FITC Chroma ET525/50m
600 60 570 630 TRITC, Texas Red, dsRed, Alexa Fluor 568 Omega 600BP60 RapidBand
705 110 650 760 Cy5, Alexa Fluor 647, Draq5    

Objectives

Magnification NA Immersion medium
5 0.16 Air
10 0.3 Air
20 0.8 Air
20 1 Water
40 1.1 Water

Detectors

Exporting a video file

This document explains how to export a video file from a time-lapse dataset.

  1. Click on a well in imaging dataset
  2. Click on the Timepoints panel, then press CTRL+A to select all the timepoints

    Harmony_SelectTimepoints.PNG

  3. Right-click on one of the selected timepoints and select "Use as Time Window"

    image1808.png

  4. Wait for the images to finish loading. Then right-click on the image and select "Save Movie ..."

    Harmony_SaveMovie.PNG


Exporting and deleting files

After your session, you should backup your data from the Opera Phenix. There are two main file formats for you to export your data: 1) as a TIFF series for analysis using MATLAB, ImageJ/Fiji, other programs, or 2) as an Archive file (compatible with Harmony only). The instructions for each are below (note that in the Harmony software “Measurements” refers to your images).

Exporting your data does not “delete” your data from Harmony. This is done through a separate process described below.

Exporting images as a TIFF series

This option should only be used for analyzing your data outside of Harmony. The TIFF series cannot be reimported back into Harmony.

To export the images:

  1. Select Settings -> Data Management -> Export Data.
  2. Select "Measurements Including Associated Files".
Exporting the archive

In the Harmony software, "archive" refers to a full and true backup of your data. Archives can be reimported back into Harmony for future analysis.

To export the data archive:

  1. Select Settings -> Data Management -> Write Archive
  2. Select the measurement files you wish to archive
  3. Select the destination path

Tips:

Opera Phenix Data Deletion Protocol

Once you have backed up/transferred your data off the Harmony server, please delete your exported data from the User Data folder and your raw data through Harmony.

To delete data using Harmony:

  1. Select Settings -> Data Management -> Delete Data
  2. Please delete your Measurements only. The other files (experiment, analysis) can remain as these do not consume much disk space.

PerkinElmer - Opera Phenix manual

This microscope has excellent online resources that you can consult after completing your in-person training.